Picomag IO-Link
Diagnostics and troubleshooting
Hauser
35
11
Diagnostics and troubleshooting
11.1 General troubleshooting
For local display
Error
Possible causes
Remedial action
Local display dark and no output
signals
Supply voltage does not match the
voltage specified on the nameplate.
Apply the correct supply voltage
→ 41.
Supply voltage has incorrect
polarity.
Correct the polarity.
The connecting cables are not
connected correctly.
Check the cable connection and
correct if necessary.
For output signals
Error
Possible causes
Remedial action
Device shows correct value on local
display, but signal output is
incorrect, though in the valid range.
Parametrization errors
Check parameterization and correct
it.
Device measures incorrectly.
Configuration error or device is
operated outside the application.
1. Check and correct parameter
configuration.
2. Observe limit values specified in
the "Technical Data".
For access
Error
Possible causes
Remedial action
Measuring device not in
smartphone or tablet live list
Bluetooth communication is
disabled
1. Check whether the Bluetooth
logo is visible on the local display or
not.
2. Re-enable Bluetooth
communication by tapping the
device three times.
Device not responding via
SmartBlue app
No Bluetooth connection
Enable Bluetooth function on
smartphone or tablet.
The device is already connected to
another smartphone/tablet.
Check whether the device is already
connected.
Login via SmartBlue app not
possible
Device is being put into operation
for the first time
Enter initial password (device serial
number) and change.
Device cannot be operated via
SmartBlue app
Incorrect password entered
Enter correct password.
Password forgotten
Contact Hauser Service.
No write access to parameters
Current user role has limited access
authorization
• Check user role
• Enter the correct customer-
specific access code → 20