Commissioning
Picomag IO-Link
26
Hauser
The current pulse repetition frequency is calculated from the current flow and the
configured pulse value:
Pulse repetition frequency = flow/pulse value
Example
• Flow: 24 l/min
• Pulse value: 0.001 l
• Pulse repetition frequency = 400 Pulse/s
The pulse output only outputs positive flow components in the set installation direction.
Negative flow components are ignored and not balanced.
Low flow cut off → 23 resets the pulse output.
Configuring the switch output
The Switch output submenu contains all the parameters that must be configured for the
configuration of the switch output.
Navigation
Menu: "Guidance" → Output 1
Menu: "Guidance" → Output 2
Parameter overview with brief description
Parameter
Description
Selection/input
Factory setting
Polarity
Select the switching
behavior.
• NPN (low-side-switch)
Switches load to low side to L-
• PNP (high-side-switch)
Switches load to high side to L+
PNP (high-side-
switch)
Switch output
function
• Off
The switch output is permanently switched off (open, non-conductive).
• On
The switch output is permanently switched on (closed, conductive).
• Diagnostic behavior
The output switches off when an event with the status signal F occurs
• Limit volume flow
Indicates if a specified limit value has been reached for the process variable.
• Limit temperature
Indicates if a specified limit value has been reached for the process variable.
• Limit conductivity
Indicates if a specified limit value has been reached for the process variable.
• Limit volume totalizer
• Range volume flow
• Range temperature
• Range conductivity
• Range volume totalizer
• Empty pipe detection
Output switches off if empty pipe detection is activated.
Off
Switch-on value
Enter the measured value
for the switch-on value.
Floating point number with sign
1
000 m³/h
Switch-off value
Enter the measured value
for the switch-off value.
Floating point number with sign
1
000 m³/h