Appendix
Picomag IO-Link
54
Hauser
Designation
ISDU
(hex)
ISDU
(dec)
Size
(byte)
Data type Access
Value range
Factory setting
Range
limits
Output 1
Operating Mode
IO-Link is set if connected to a
master
0x01F4
500
2
uint
r/w
P-Out=0
I-Out=1
S-In=2
S-Out=3
IO-Link=4
U-Out=5
off=6
IO-Link
Current output I-Out 1
I - OUT Assign
0x0258
600
2
uint
r/w
off=0
volume flow=1
temperature=2
conductivity=4
volume flow
Q-Start-Value
ASP
Unit selection list from
Unit
Volumeflow
0x0259
601
4
float
r/w
0 l/min
-9.9·10⁹
9.9·10⁹
Q-End-Value
AEP
Unit selection list from
Unit
Volumeflow
0x025A
602
4
float
r/w
25/50/100/750
l/min
-9.9·10⁹
9.9·10⁹
T-Start-Value
ASP
Unit selection list from
Unit
Temperature
0x025F
607
4
float
r/w
-10 °C
-9.9·10⁹
9.9·10⁹
T-End-Value
AEP
Unit selection list from
Unit
Temperature
0x0260
608
4
float
r/w
+70 °C
-9.9·10⁹
9.9·10⁹
s-Start-Value
ASP
Unit selection list from
Unit
Conductivity
0x025D
605
4
float
r/w
0
-9.9·10⁹
9.9·10⁹
s-End-Value
AEP
Unit selection list from
Unit
Conductivity
0x025E
606
4
float
r/w
1000
-9.9·10⁹
9.9·10⁹