Model OUM980 with OUSAF21 Inline Sensor
Operation
17
Table 4.1 shows the status display description and display condition.
Status Area Name
Displayed symbol
Description and Display criteria
Analyzer Label
“xxxxxxxxxxxx”
Editable up to 12 characters for user
analyzer identification.
“OD”
Display “OD” under normal condition,
Process Unit
“xxxxxx”
Displays user configured engineering unit
if PV correlation function is activated.
Over Range Symbol
“OVER”
Displays if measurable range is reached.
“BL OFF”
Displays if BaseLine is not activated.
“BL ON”
Displays if BaseLine function is activated.
BaseLine Status
“BL SFT”
Displays if BaseLine and BaseLine Shift
function are both activated.
Alarm
“AL x”
Displays if corresponding relays (1, 2, or
both) are triggered.
“LAMP LOW”
Flash if reference voltage is lower than
80% of initial value.
“LAMP FAIL”
Displays if lamp reference signal is lost.
Lamp Status
“LAMP OFF”
Displays if lamp is turned off.
Cable Status
“CABLE FAULT”
Displays if lamp or signal cable has a faulty
connection.
Table 4.1: Operation screen under measuring mode
Figure 4.4: Operations under Measuring Mode
Under measuring mode, press
button will let you enter the process control screen, where user can operate process
control routine. The detail of process control function will be listed in Chapter 6.