Model OUM980 with OUSAF21 Inline Sensor
Commissioning
25
Easycal
TM
Installation
Enter SYSTEM DATA sub-menu to access
CHANGE EASYCAL CNFG function.
Press
to continue.
Use
to select [YES] or [NO] to decide if the
Easycal
™
is installed. To install, select [YES] and confirm.
This screen displays the present NIST Filter Absorbance
values and the re-certification date. Press
to change
the date and filter values starting with the next screen.
Move cursor to edit the day, month and year. Save data
and move to next step.
Refer NIST filter high absorption value listed on NIST filter
certificate delivered with Easycal
™
. The value should be
close to 1.0000 AU. Press
to continue and use the
same procedure to input the low filter absorption value,
close to 0.5000 AU.
After put in both filter data use
to confirm the
setting. Easycal
™
data installation is finished.