Model OUM980 with OUSAF21 Inline Sensor
Commissioning
30
5.3.6
Current Output Calibration
This function calibrates 4~20mA current output to assure the measurement accuracy.
Calibrate 4~20mA Output
Enter MAINTENANCE Sub-Menu to access
OUTPUT CAL function. Move cursor to OUTPUT 1 and press
to
select this channel.
Move the cursor up and down to select the value need to be trimmed.
Use
to trim 4mA or 20mA.
The trimming resolution is 1% per step. Press
to accept and
confirm the trimming.
Follow same procedure to trim OUTPUT 2.
5.3.7
Date and Time Setting
Set Clock
Enter MAINTENANCE sub-menu to access SET CLOCK function. Use
to select the CUR DATE line.
Press
to select the DATE option and then the CHANGE DATE
Screen is displayed with the current date. Use
to position
the cursor and
to change the mm/dd/yy numbers. When
the wanted date is entered, press
to accept and confirm.
Press
to select the TIME option and then the CHANGE TIME
Screen is displayed with the current time. Use
to position
the cursor and
to change the hh/mm/ss numbers. When
the wanted time is entered, press
to accept and confirm.