Model OUM980 with OUSAF21 Inline Sensor
Commissioning
27
Pathlength Setting
Enter SYSTEM DATA sub-menu to access
PATHLENGTH function. Press
to continue.
The FROM value is the present pathlength Value. Use
to
position the cursor and
to select wanted digit. Repeat for
each digit position. When new pathlength has been selected, press
to confirm the change.
The SYS DATA: PATHLENGTH is displayed together with the new
values for pathlength and OD MIN/MAX.
5.3.4
Setup Output Span
The 4~20mA span setting will provide user capability to achieve maximum resolution for preferred measurement range.
Change 4~20mA Output Span
Enter PROCESS SET-UP sub-menu to access OUTPUT SPANS
function. Use
to move the cursor to O1 and press
to select.
This screen will display the current output span MIN/MAX span based
on pathlength or process value correlation setting. Press
to
change the span.
In this screen the current value for span is the FROM value. Position
the cursor as done previously and select the number for that position.
Repeat as needed for each digit position. When the new span has been
entered
,
press
and confirm the sanity check to accept.
Use same procedure above to change Channel 2 output span.