Model OUM980 with OUSAF21 Inline Sensor
Technical Information
43
8.
Technical Information
8.1
Photometric Analyzer OUM980
8.1.1
Input
Measured Variable
Dual Channel Current from Model OUSAF21 Inline Sensor
Range
5 decades of absorbance signal from each detector
Digital Input
4 Channels, dry contact
Keyboard
6 push buttons keyboard
8.1.2
Output
Current Output
2 channel, 4~20mA NAMUR Compatible
Resolution
12bit
O1 to PE
500 VDC
O2 to PE
500 VDC
Isolation
O1 to O2
500 VDC
Signal on Lamp Alarm
3.6mA or 21.5mA in case of error
Actual Current
Output Range
3.8mA to 20.5mA
Load
750 ohm Max.
Front Panel Display
4x20 Character LCD display
Alarm LED
On front Panel, Red LED
General Relay Output
2 Channel, Solid State Relays
Relay Operation
Above/Below trigger configurable; NO/NC configurable
Relay Delay
0~60 sec configurable
Lamp Alarm Relay
1 Channel, NC
Lamp Power Supply
3 to 5 vdc, 1.5A Max.
8.1.3
Power Supply
Supply Voltage
85~265 VAC, 24VDC as option
Power Consumption
12 VA Max.
Mains Fuse
3.15 A