Model OUM980 with OUSAF21 Inline Sensor
Commissioning
24
5.3.2.2
Filter calibration Set up and Calibration
The
Easycal
TM
unit contains an optical that was selected based on laboratory measurement of the customers submitted sample.
This filter is placed into the optical measurement path of the instrument and displays an absorbance ratio process related. This filter
has been scanned with a traceable source and actual absorbance at individual (measurement) wavelengths is ascertained.
The
Easycal
TM
allows the calibration without using liquid standards. Refer to the actual value of the
Easycal
TM
optical filter as
noted on the Calibration Certificate supplied with the unit. The absorbance value should be entered into the Model OUM980 as
part of the configuration setup.
Figure 5.2: Easycal
TM
Control for the OUSAF22 Dual Beam Sensor.