Model OUM980 with OUSAF21 Inline Sensor
Commissioning
22
5.3.1
Factory Default Setting
Parameters
Factory Default Value
Measurement Unit
OD
Measurement Pathlength
100mm to 250mm
Output Span 1
5.0000 OD
Output Span 2
5.0000 OD
Relay 1 Set Point/Status
1.0000 OD/ Disabled
Relay 2 Set Point/Status
1.0000 OD/ Disabled
Process Value Correlation
Off (Curve #0)
BaseLine
Disabled
Password
Disabled
Table 5.1: Factory Default Setting
5.3.2
Sensor Calibration
Model OUM980 analyzer supports 2 different calibration procedures: Liquid calibration and Filter Calibration.
5.3.2.1
Liquid Calibration Procedure
A liquid solution of a known optical density (at the wavelength of the sensor) must be used for instrument calibration. The
following procedure is used in factory. Calibration liquid can be chosen by user depending on application.