Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Description of parameters
Hauser
51
P01-xxxxxxxx-19-xx-xx-xx-082
Fig. 25:
Quick Setup menu, "Flow" measuring mode
FULL CALIB. (315)/(004)
Entry
Enter level value for the upper calibration point (container full).
The container is either completely or almost full. By entering a value for this parameter,
you are assigning a level value to the pressure present at the device.
Prerequisite:
• LEVEL SELECTION = Level Easy Pressure (
see also Page 46),
CALIBRATION MODE = Wet (
see also Page 60)
• LEVEL SELECTION = Level Standard,
LEVEL MODE = Linear (
see also Page 68),
CALIBRATION MODE = Wet (
see also Page 73)
!
Note!
For this parameter, the on-site display shows the level value to be entered and the
pressure present at the device. In order for the level value to be saved together with the
pressure present at the device, the entry field for the level value must first be activated
using the "+"- or "–" key before confirming with the "E" key. This applies also if the level
value is to remain unchanged.
Factory setting:
100.0
DAMPING VALUE (247)
Entry
Enter damping time (time constant
).
The damping affects the speed at which all subsequent elements, such as the on-site
display, measured value and current output react to a change in the pressure.
Input range:
0.0...999.0 s
Factory setting:
2.0 s or as per order specifications
Table 4: (GROUP SELECTION
)
QUICK SETUP "Level"
Parameter name
Description
1)
1)
2)
1)
M
AX
. FL
O
W
M
AX
. PRE
SS
. FL
O
W
P
OS
. ZER
O A
DJU
ST
D
A
MPIN
G
V
A
LUE
Flow
Pressure
Level
O
PER
AT
IN
G
MENU
M
e
asu
red
valu
e
G
R
O
UP
S
ELEC
T
I
O
N
L
A
N
G
U
AG
E
ME
AS
URIN
G
M
O
DE
QUICK
S
E
T
UP
ME
AS
URIN
G
M
O
DE
1) Display via
o
n-sit
e d
isplay
o
nly
2) Display via
d
igital
co
mmuni
c
ati
o
n