Level measurement
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
20
Hauser
Description
1
Deltabar S: Before you configure the device for your
application, the pressure piping must be cleaned and
filled with medium. See Operating Instructions
BA00270P, Section 6.5.1
P01-PMC71xxx-19-xx-xx-xx-001
Fig. 7:
Calibration with reference pressure –
wet calibration
1
See Table, Steps 10 and 11.
2
See Table, Step 12.
3
See Table, Step 13.
2
Carry out position adjustment if necessary. See Page 53,
Table 6: POSITION ADJUSTMENT.
3
Select the "Level" measuring mode via the MEASURING
MODE parameter.
On-site display:
Menu path: GROUP SELECTION
MEASURING
MODE
Digital communication:
Menu path: OPERATING MENU
SETTINGS
BASIC SETUP
MEASURING MODE
4
If necessary, select the "Level Easy Height" level mode
using the LEVEL SELECTION parameter.
On-site display:
Menu path: GROUP SELECTION
MEASURING
MODE "Level"
LEVEL SELECTION
Digital communication:
Menu path: OPERATING MENU
SETTINGS
BASIC SETUP
MEASURING MODE "Level"
LEVEL SELECTION
5
On-site display:
Select BASIC SETUP function group. Menu path:
GROUP SELECTION
OPERATING MENU
SETTINGS
BASIC SETUP
1000 l
0,5 m
4,5 m
0 l
ρ
= 1
kg
dm
3
1
3
2