Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Trouble-shooting
Hauser
135
704 (A704)
Alarm
B
Funktion
check (C)
B>Measurement error
– Fault in the main electronics.
– Briefly disconnect device from
the power supply.
12
– Main electronics defect.
– Replace main electronics.
705 (A705)
Alarm
B
Failure (F)
B>Measurement error
– Fault in the main electronics.
– Briefly disconnect device from
the power supply.
21
– Main electronics defect.
– Replace main electronics.
706 (W706)
Warning
C
Maintenance
request (M)
C>Configuration in HistoROM
and device not identical
– Configuration (parameters) in the
HistoROM and in the device is
not identical.
– Copy data from the device to the
HistoROM.
– Copy data from the HistoROM to
the device. The message remains
if the HistoROM and the device
have different software versions.
The message goes out if you copy
the data from the device to the
HistoROM.
– Device reset codes such as 7864
do not have any effect on the
HistoROM. That means that if
you do a reset, the configurations
in the HistoROM and in the
device may not be the same.
See also Operating Instructions
BA00270P (Deltabar S),
BA00271P (Cerabar S),
BA00332P (Deltapilot S) Section
5.5.1 "Copying configuration data".
59
707 (A707)
Alarm
B
Funktion
check (C)
B>X-VAL. of lin. table out of
edit limits.
– At least one X-VALUE in the
linearisation table is either below
the value for HYDR. PRESS MIN.
or MIN. LEVEL or above the
value for HYDR. PRESS. MAX. or
LEVEL MAX.
– Carry out calibration again. (
See also these Operating
Instructions, chapter 5.)
38
710 (W710)
Warning
C
Funktion
check (C)
B>Set span too small. Not
allowed.
– Values for calibration (e.g. lower
range value and upper range
value) are too close together.
– Adjust calibration to suit sensor.
(
See also Page 119, parameter
description MINIMUM SPAN.)
51
– The sensor was replaced and the
customer-specific configuration
does not suit the sensor.
– Adjust calibration to suit sensor.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried out. – Check configuration and perform
download again.
711 (A711)
Alarm
B
Funktion
check (C)
B>LRV or URV out of edit
limits
– Lower range value and/or upper
range value undershoot or
overshoot the sensor range limits.
– Reconfigure lower range value
and/or upper range value to suit
the sensor. Pay attention to
position factor.
37
– The sensor was replaced and the
customer-specific configuration
does not suit the sensor.
– Reconfigure lower range value
and/or upper range value to suit
the sensor. Pay attention to
position factor.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried out. – Check configuration and perform
download again.
713 (A713)
Alarm
B
Funktion
check (C)
B>100% POINT level out of
edit limits
– The sensor was replaced.
– Carry out calibration again.
39
Code
Error type/
NA 64
Corresponds
NE 107
Message/description
Cause
Measure
Prio
rity