Description of parameters
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
60
Hauser
OUTPUT UNIT (023)
Selection
Select unit for measured value display and MEASURED VALUE parameter (
Page 122).
!
Note!
The selected unit is used only to describe the measured value. This means that when
selecting a new output unit, the measured value is not converted.
Example:
• current measured value: 0.3 ft
• new output unit: m
• new measured value: 0.3 m (9.8 ft)
Options
• %
• mm, cm, dm, m
• ft, inch
• cm
3
, dm
3
, m
3
,m
3
E
3
• l, hl
• ft
3
, ft
3
E
3
• gal, bbl, Igal
• g, kg, t
• lb, ton, oz
Factory setting:
%
CALIBRATION MODE (008)
Selection
Select calibration mode.
Options:
• Wet
Wet calibration takes place by filling and emptying the container. In the case of two
different levels, the level, volume, mass or percentage value entered is assigned to the
pressure measured at this point in time. (
See also this table, parameter descriptions
for EMPTY CALIB. and FULL CALIB.)
• Dry
Dry calibration is a theoretical calibration. For this calibration, specify two pressure-
level value pairs via the following parameters: EMPTY CALIB., EMPTY PRESSURE,
FULL CALIB. and FULL PRESSURE.
Factory setting:
Wet
EMPTY CALIB. (010)
Entry
Enter level, volume, mass or percentage value for the lower calibration point (container
empty).
The container is either empty or part full. By entering a value for this parameter, you
assign a level, volume, mass or percentage value to the pressure present at the device.
The unit is selected via the OUTPUT UNIT parameter (
Prerequisite:
• CALIBRATION MODE = Wet
!
Note!
For this parameter, the on-site display shows the level value to be entered and the
pressure present at the device. In order for the level value to be saved together with the
pressure present at the device, the entry field for the level value must first be activated
using the "+"- or "–" key before confirming with the "E" key. This applies also if the level
value is to remain unchanged.
Factory setting:
0.0
FULL CALIB. (004)
Entry
Enter height, volume or mass value for the upper calibration point (container full).
The container is either completely or almost full. By entering a value for this parameter,
you assign a height, volume or mass value to the pressure present at the device. The unit
is selected via the OUTPUT UNIT parameter (
Prerequisite:
• CALIBRATION MODE = Wet
!
Note!
For this parameter, the on-site display shows the level value to be entered and the
pressure present at the device. In order for the level value to be saved together with the
pressure present at the device, the entry field for the level value must first be activated
using the "+"- or "–" key before confirming with the "E" key. This applies also if the level
value is to remain unchanged.
Factory setting:
100.0
Table 8: (GROUP SELECTION
) OPERATING MENU
SETTINGS
BASIC SETUP "Level",
LEVEL SELECTION "Level Easy Pressure"