Figure A-13: Ambient Temperature vs. Process Temperature
A
B
185 (85)
131 (55)
100 (38)
50 (10)
-40 (-40)
392 (200)
-17 (-27)
-40 (-40)
200 (93)
400 (204)
600 (316)
800
(427)
-320
(-196)
752 (400)
A. Ambient temperature °F (°C)
B. Process temperature °F (°C)
Figure A-14: Ambient Temperature vs. Process Temperature with Option Code BR5
A
B
185 (85)
131 (55)
100 (38)
50 (10)
-67 (-55)
392 (200)
-17 (-27)
-67 (-55)
200 (93)
400 (204)
600 (316)
800
(427)
-320
(-196)
752 (400)
A. Ambient temperature °F (°C)
B. Process temperature °F (°C)
Note
Nozzle insulation for the HTHP version (Operating Temperature and Pressure code H)
should not exceed 4 in. (10 cm) of height above the flange.
Note
In applications where the ambient temperature exceeds the limits of the electronics, a
Remote Mounting connection can be used. The maximum temperature for the Remote
Mounting connection at the vessel connection point is 302 °F (150 °C).
Table A-10: Ambient Temperature Limits
Description
Operating limit
Storage limit
Without integral display
-40 °F to 185 °F (-40 °C to 85 °C) -58 °F to 194 °F (-50 °C to 90 °C)
With integral display
-40 °F to 158 °F (-40 °C to 70
°C)
-40 °F to 185 °F (-40 °C to 85 °C)
Option code BR5 without
integral display
-67 °F to 185 °F (-55 °C to 85 °C) -76 °F to 194 °F (-60 °C to 90 °C)
Specifications and reference data
Reference Manual
April 2019
00809-0100-4530
292
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...