Tools required
•
HART host/communicator
•
Current meter
8.7.2
Prepare for proof-testing
Prepare for proof-testing using Rosemount Radar Master
Note that prior to these tests, inspect the echo curve to ensure that no disturbing echoes
affecting the measurement performance are present.
Procedure
Select
Setup
→
Echo Curve
.
Prepare for proof-testing using AMS Device Manager and
handheld communicator
Note that prior to these tests, inspect the echo curve to ensure that no disturbing echoes
affecting the measurement performance are present.
Procedure
Select
Service Tools
→
Echo Tuning
→
Echo Curve
.
8.7.3
Suggested comprehensive proof-test
Suggested comprehensive proof-test with Rosemount Radar
Master
The suggested proof-test described below will detect 94% of possible DU failures in the
Rosemount 5300 Level Transmitters.
Procedure
1. Bypass the safety function and take appropriate action to avoid a false trip.
2. Disable write protection in device (if enabled).
a) Select
Tools
→
Lock/Unlock Configuration Area
.
b) Enter password to unlock.
3. Using Loop Test, enter current value (mA) representing high alarm current. Verify
that analog output current is correct using the reference meter.
a) Select
Setup
→
Output
→
Analog Out 1
and click
Loop test
.
b) Enter current value representing high alarm current.
c) Click
Start
to output current.
d) Verify that analog output current is correct.
Safety Instrumented Systems (4-20 mA only)
Reference Manual
April 2019
00809-0100-4530
252
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...