Table I-1: Resource Transducer Block Parameters
(continued)
Parameter
Index number
Description
TAG_DESC
02
The user description of the intended application of the block.
TEST_RW
08
Read/write test parameter - used only for conformance testing.
UPDATE_EVT
35
This alert is generated by any change to the static data.
WRITE_ALM
40
This alert is generated if the write lock parameter is cleared.
WRITE_LOCK
34
When hardware write protection is selected, WRITE_LOCK becomes an
indicator of the jumper setting and is unavailable for software write
protection.
When software write lock is selected, and WRITE_LOCK is set, no
writings from anywhere else are allowed, except to clear WRITE_LOCK.
Block input will continue to be updated.
WRITE_PRI
39
Priority of the alarm generated by clearing the write lock.
XD_OPTION
44
Indicates which transducer block licensing options are enabled.
I.3
Alerts
The Resource Block will act as a coordinator for alerts. There will be three alarm
parameters (FAILED_ALARM, MAINT_ALARM, and ADVISE_ALARM) which will contain
information regarding some of the device errors which are detected by the transmitter
software. There will be a RECOMMENDED_ACTION parameter which will be used to display
the recommended action text for the highest priority alarm, and a HEALTH_INDEX
parameter (0 - 100) indicating the overall health of the transmitter. FAILED_ALARM will
have the highest priority followed by MAINT_ALARM, and ADVISE_ALARM will be the
lowest priority.
I.3.1
FAILED_ALARMS
A failure alarm indicates a failure within a device that will make the device or some part of
the device non-operational. This implies that the device is in need of repair and must be
fixed immediately. There are five parameters associated with FAILED_ALARMS specifically,
they are described below.
FAILED_ENABLED
This parameter contains a list of failures in the device which makes the device non-
operational that will cause an alert to be sent. Below is a list of the failures:
1. Level / Interface Measurement Failure
2. Volume / Temperature Measurement Failure
3. Electronics Failure / Transducer Block
4. Probe Missing
5. Input / Output Failure
6. Non-Volatile Memory Failure
Resource Transducer Block
Reference Manual
April 2019
00809-0100-4530
450
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...