5.2.1
Basic configuration
The basic configuration includes parameters for a standard configuration which is
sufficient in most cases. The basic configuration comprises the following items:
•
Measurement units
•
Tank configuration - Tank geometry - Environment - Volume
•
Analog output
5.2.2
Echo tuning
Amplitude thresholds can be adjusted in order to handle special situations when, for
example, objects in the tank cause disturbing echoes that are stronger than the surface
echo. A useful function is the Amplitude Threshold Curve (ATC) which lets you filter out
single disturbing echoes. See
5.2.3
LCD configuration
It is possible to specify the variables to be presented on the display panel. See also
.
5.2.4
Advanced configuration
For some applications, further device specific configuration is needed in addition to the
basic configuration. This may be due to product properties or tank shape. See
for more information.
5.2.5
Configuration tools
The transmitter can be configured using:
•
Rosemount Radar Master (RRM). Note that RRM is recommended for advanced
configuration features. See
Basic configuration using Rosemount Radar Master
information on how to use RRM for configuration of the Rosemount 5300 Level
Transmitter.
•
Device Descriptor (DD) based systems, e.g. AMS Device Manager, 475 Field
Communicator, AMS Trex
™
Device Communicator, and DeltaV
™
, or any other EDDL or
enhanced-EDDL host. See
Basic configuration using AMS Device Manager and
Basic configuration using DeltaV
(only for F
OUNDATION
™
Fieldbus), for information on how to use DDs for configuration of the Rosemount 5300
Level Transmitter.
See
for AMS Device Manager and handheld communicator menu tree
structure.
RRM is a user-friendly, Windows
™
-based software package including waveform plots, off-
line/on-line configuration Wizard, logging, and extensive on-line help.
To communicate with the transmitter using RRM, a HART
®
modem (part number
03300-7004-0001 or 03300-7004-0002) or a F
OUNDATION
Fieldbus modem (part number
03095-5108-0001 for PCMCIA) is required. For F
OUNDATION
Fieldbus communication you
Configuration
Reference Manual
April 2019
00809-0100-4530
114
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...