When using remote housing, make sure it is fully assembled before calibrating the
Vapor Compensation function.
Procedure
1. Verify the basic configuration for device has been performed.
2. Make sure the correct probe type is selected.
a) HART
®
: Select
Configure
→
Manual Setup
→
Level Setup
.
F
OUNDATION
™
Fieldbus: Select
Configure
→
Manual Setup
.
b) Select the
Probe
tab.
c) In the
Probe type
list, select
Coaxial HTHP
for probe type 3V or
Rigid Single
HTHP 0.3
″
(8mm)
for probe type 4U.
d) Select
Send
.
3. Check that Mounting Type is set to Pipe/Chamber and select inner diameter.
a) HART: Select
Configure
→
Manual Setup
→
Level Setup
.
F
OUNDATION
Fieldbus: Select
Configure
→
Manual Setup
.
b) Select the
Geometry
tab.
c) In the
Mounting Type
list, select
Pipe/Chamber
.
d) In the
Inner Diameter, Pipe/Chamber/Nozzle
list, select the corresponding
inner diameter.
e) Select
Send
.
4. HART: Select
Configure
→
Manual Setup
→
Level Setup
.
F
OUNDATION
Fieldbus: Select
Configure
→
Manual Setup
→
Advanced
.
5. HART: Select the
Vapor Compensation
tab.
F
OUNDATION
Fieldbus: Select the
Vapor Compensation Setup
button.
6. In the
Reference Reflector Type
list, select reference reflector as appropriate, and
then select
Next
.
Option
Description
Coax, l=350 mm (RR B3 in old DD
versions)
Use this option for probe type 3V with
short reflector
Coax, l=500 mm (RR B4 in old DD
versions)
Use this option for probe type 3V with
long reflector
d=13 mm, l=350 mm (RR A6 in old
DD versions)
Use this option for probe type 4U with
short reflector
d=13 mm, l=500 mm (RR B2 in old DD
versions)
Use this option for probe type 4U with
long reflector
7. HART: Select the
Use Vapor Compensation
checkbox, and then select
Send
.
Advanced configuration
Reference Manual
April 2019
00809-0100-4530
412
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...