L, desired probe length:
/
2. Determine n, the number of middle segments needed for the desired probe length.
See
n, number of middle segments:
3. Calculate Y, the length of the bottom segment. See
.
Y, length of bottom segment:
<
4. Continue as follows:
Length of bottom segment (Y)
Action
Y < 0.4 in. (10 mm)
<
•
Continue with step 7. Do not use the
bottom segment.
Y ≥ 0.4 in. (10 mm)
<
•
Continue with step 5 and cut the bottom
segment.
Y = 31.5 in. (800 mm)
<
a. Add one extra middle segment to the
calculated n.
b. Continue with step 7.
5. Mark where to cut the bottom segment.
1
2
3
4
5
6
7
8
0
1
2
3
0
<
6. Cut the bottom segment at the mark.
Note
Make sure the bottom segment is fixed while cutting.
Mechanical installation
Reference Manual
April 2019
00809-0100-4530
86
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...