2. Choose the tank material if recommended.
3. Proceed to
if no configuration is needed.
Trim Near Zone is described further in
.
Probe End Projection is described further in
Vapor compensation is described further in
Restart the device
When the transmitter is configured, it should be restarted to make sure that all
configuration changes are properly activated and the transmitter performs as expected.
Procedure
In the
Guided Setup
window, select
Restart the Device
.
It may take up to 60 seconds until measurement values are updated.
Verify level
The Verify Level tool matches the product level reported by the device to a reference
measurement (measured by using for example handgauging).
Reference Manual
Configuration
00809-0100-4530
April 2019
Reference Manual
135
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...