Table A-10: Ambient Temperature Limits
(continued)
Description
Operating limit
Storage limit
Option code BR5 with integral
display
-67 °F to 158 °F (-55 °C to 70
°C)
-76 °F to 185 °F (-60 °C to 85 °C)
(1)
Integral display may not be readable and device display updates will be slower at temperatures
below -4 °F (-20 °C).
A.2.10
Flange rating
The flanges, except the Fisher
™
and Masoneilan
™
flanges, are triple certified for the
materials 316, 316L, and EN 1.4404.
ASME flange rating
316 up to Class 1500 flanges according to ASME B16.5 Table 2-2.2 and 316L for Class 2500
flanges according to ASME B16.5 Table 2-2.3:
•
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
•
HP: Class 2500 up to max 482 °F (250 °C)
•
C: Class 2500 up to max 392 °F (200 °C)
•
HTHP: Class 2500 up to max 752 °F (400 °C)
Alloy C-276 (UNS N10276) according to ASME B16.5 Table 2-3.8:
•
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
•
HP: Class 1500 up to max 482 °F (250 °C)
•
HTHP: Class 1500 up to max 752 °F (400 °C)
Duplex 2205 (UNS S31803) according to ASME B16.5 Table 2-2.8:
•
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
•
HP: Class 1500, -51 °F (-46 °C) up to max 482 °F (250 °C)
•
HTHP: Class 1500, -51 °F (-46 °C) up to max 599 °F (315 °C)
EN flange rating
EN 1.4404 according to EN 1092-1 material group 13E0:
•
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
•
HP: PN 320 up to max 482 °F (250 °C)
•
C: PN 320 up to max 392 °F (200 °C)
•
HTHP: PN 320 up to max 752 °F (400 °C)
Alloy C-276 (UNS N10276) according to EN 1092-1 material group 12E0:
•
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
•
HP: PN 320 up to max 482 °F (250 °C)
•
HTHP: PN 320 up to max 752 °F (400 °C)
Reference Manual
Specifications and reference data
00809-0100-4530
April 2019
Reference Manual
293
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...