5.4.3 Procedure: Configure Outputs
Under the Program/Outputs menu, the adjacent screen
will appear to allow configuration of the outputs. Follow
the menu screens in Fig. 5-2 to configure the outputs.
5.4.4 Procedure: Assigning Measurements the Low
and High Current Outputs
The adjacent screen will appear when entering the
Assign function under Program/Output/Configure.
These screens allow you to assign a measurement,
process value, or temperature input to each output.
Follow the menu screens in Fig. 5-2 to assign
measurements to the outputs.
5.4.5 Procedure: Ranging the Current Outputs
The adjacent screen will appear under
Program/Output/Range. Enter a value for 4mA and
20mA (or 0mA and 20mA) for each output. Follow the
menu screens in Fig. 5-2 to assign values to the out-
puts.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Output
M
Configure
Assign: S1 Meas
Range: 4-20mA
Scale: Linear
Dampening: 0sec
Fault Mode: Fixed
Fault Value: 21.00mA
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Output
M
Assign
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Output Range
OM SN 4mA: 0.000µS/cm
OM SN 20mA: 20.00µS/cm
OM SN 4mA: 00.00pH
OM SN 20mA: 14.00pH
Figure 5-2. Configuring and Ranging the Current Outputs
Instruction Manual
LIQ-MAN-1056
Section 5.0: Programming the Analyzer - Basics
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...