91
7.6.3.2 zeroing The Sensor
The adjacent screen will appear during Zero Cal. Be
sure sensor has been running in zero solution for at
least two hours before starting zero step.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Zeroing
Wait
The adjacent screen may appear if In Zero Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
The adjacent screen will appear if In Zero Cal is suc-
cessful. The screen will return to the Amperometric
Cal Menu.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Sensor zero failed
Press EXIT
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Sensor zero done
If the In Process Cal is successful, the screen will
return to the Cal sub-menu.
The adjacent screen may appear if In Process Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
7.6.3.3 In Process Calibration
The adjacent screen will appear prior to In Process Cal
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Calibration
Error
Press EXIT
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Wait for stable
reading.
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...