Model 1
µ
S/cm 10
µ
S/cm 100
µ
S/cm 1000
µ
S/cm 10mS/cm 100mS/cm 1000mS/cm 2000mS/cm
5
µ
S/cm to 500mS/cm
15
µ
S/cm to 1500mS/cm
500mS/cm to 2000mS/cm
500
µ
S/cm to 2000mS/cm
100
µ
S/cm to 2000mS/cm
1500mS/cm to 2000mS/cm
226
242
222
(1in & 2in)
225 & 228
Measures conductivity in the range of 1 (one)
µ
S/cm to
2,000,000
µ
S/cm (2 S/cm), Measurement choices are
conductivity, resistivity, total dissolved solids, salinity,
and % concentration. The % concentration selection
includes the choice of five common solutions (0-12%
NaOH, 0-15% HCl, 0-20% NaCl, and 0-25% or
96-99.7% H
2
SO
4
). The conductivity concentration
algorithms for these solutions are fully temperature
compensated. For other solutions, a simple-to-use
menu allows the customer to enter his own data. The
analyzer accepts as many as five data points and fits
either a linear (two points) or a quadratic function (three
or more points) to the data. Two temperature compensation
options are available: manual slope (X%/°C) and neutral
salt (dilute sodium chloride). Temperature compensation
can be disabled, allowing the analyzer to display raw
conductivity. Reference temperature and linear temper-
ature slope may also be adjusted for optimum results.
For more information concerning the use and operation
of the toroidal conductivity sensors, refer to the product
data sheets.
Repeatability:
±0.25% ±5
µ
S/cm after zero cal
Input filter:
time constant 1 - 999 sec, default 2 sec.
Response time:
3 seconds to 100% of final reading
Salinity:
uses Practical Salinity Scale
Total Dissolved Solids:
Calculated by multiplying
conductivity at 25ºC by 0.65
Temperature Specifications:
Recommended Sensors
All Rosemount submersion/immersion and flow-
through toroidal sensors.
1.4 Toroidal Conductivity (Code -21 and -31)
Temperature range
-25 to 210ºC (-13 to 410ºF)
Temperature Accuracy,
Pt-100, -25 to 50 ºC
± 0.5ºC
Temperature Accuracy,
Pt-100,. 50 to 210ºC
± 1ºC
Performance Specifications
Recommended Range - Toroidal Conductivity
Model 226: ±1% of reading ±5
µ
S/cm in recommended range
Models 225 & 228: ±1% of reading ±10
µ
S/cm in recommended range
Models 222,242: ±4% of reading in recommended range
Model 225, 226 & 228: ±5% of reading outside high recommended range
Model 226: ±5
µ
S/cm outside low recommended range
Models 225 & 228: ±15
µ
S/cm outside low recommended range
High performance toroidal conductivity sensors
Models 226 and 225
LOOP PERFORMANCE (Following Calibration)
5
Instruction Manual
LIQ-MAN-1056
Section 1.0: Description and Specification
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...