1.9 Dissolved Oxygen (Code -25
and -35)
The 1056 is compatible with the 499ADO, 499ATrDO,
Hx438, and Gx438 dissolved oxygen sensors and the
4000 percent oxygen gas sensor. The 1056 displays
dissolved oxygen in ppm, mg/L, ppb,
µ
g/L, % satura-
tion, % O
2
in gas, ppm O
2
in gas. The analyzer fully
compensates oxygen readings for changes in mem-
brane permeability caused by temperature changes.
An atmospheric pressure sensor is included on all dis-
solved oxygen signal boards to allow automatic atmos-
pheric pressure determination at the time of calibration.
If removing the sensor from the process liquid is imprac-
tical, the analyzer can be calibrated against a standard
instrument. Calibration can be corrected for process
salinity. For more information on the use of ampero-
metric oxygen sensors, refer to the product data
sheets.
Performance Specifications
Resolution:
0.01 ppm; 0.1 ppb for 499A TrDO sensor
(when O
2
<1.00 ppm); 0.1%
Input Range:
0 nA – 100
µ
A
Temperature Compensation:
Automatic (via RTD) or
manual (0-50 °C).
Input filter:
time constant 1 - 999 sec, default 5 sec.
Response time:
6 seconds to 100% of final reading
Recommended Sensors
Rosemount
amperometric
membrane
and
steam-sterilizable sensors listed above
1.10 Dissolved Ozone (Code -26
and -36)
The 1056 is compatible with the Model 499AOZ sen-
sor. The 1056 fully compensates ozone readings for
changes in membrane permeability caused by temper-
ature changes. For more information concerning the
use and operation of the amperometric ozone sensors,
refer to the product data sheets.
Performance Specifications
Resolution:
0.001 ppm or 0.01 ppm – selectable
Input Range:
0nA – 100
µ
A
Temperature Compensation:
Automatic (via RTD) or
manual (0-35°C)
Input filter:
time constant 1 - 999 sec, default 5 sec.
Response time:
6 seconds to 100% of final reading
Recommended Sensor
Rosemount Model 499A OZ ozone sensor
Dissolved Oxygen sensor with Variopol connection
Model 499ADO
Dissolved Ozone sensors with Polysulfone body
Variopol connection and cable connection
Model 499AOZ
9
Instruction Manual
LIQ-MAN-1056
Section 1.0: Description and Specification
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...