6.2.4 Solution Temperature Correction
The display screen for selecting the Solution
temperature correction algorithm is shown. The default
value is displayed in
bold type
. Refer to the pH/ORP
Programming flow diagram to complete this function.
6.2.5 Temperature Coefficient
The display screen for entering the custom solution tem-
perature coefficient is shown. The default value is dis-
played in
bold type.
Refer to the pH/ORP
Programming flow diagram to complete this function.
6.2.6 Resolution
The display screen for selecting 0.01pH or 0.1pH for pH
display resolution is shown. The default value is displayed
in
bold type
. Refer to the pH/ORP Programming flow
diagram to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Sol’n Temp Corr.
Off
Ultra Pure Water
High pH
Custom
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Resolution
0.01pH
0.1pH
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Sol’n Temp Coeff.
-
0.032pH/ºC
6.2.7 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type.
Refer to the pH/ORP Programming flow diagram
to complete this function.
6.2.8 Reference Impedence
The display screen for selecting Low or High Reference
impedance is shown. The default value is displayed in
bold type
. Refer to the pH/ORP Programming flow diagram
to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Reference Z
Low
High
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Input filter
04 sec
6.3 ORP Measurement Programming
6.3.1 Description
The section describes how to configure the Model 1056 analyzer for ORP measurements. The following programming
and configuration functions are covered:
Measure
Sec.
Menu function:
default
Description
ORP
6.3.2
Measurement type: pH
Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
6.3.3
Preamp location: Analyzer
Identify preamp location
6.3.4
Filter: 4 sec
Override the default input filter, enter 0-999 seconds
6.3.5
Reference Z: Low
Select low or high reference impedance
TABLE 6-2. ORP Measurement Programming
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...