102
The following screen may appear if Grab Cal is
unsuccessful.
The following screen will appear if Grab Cal is successful.
The screen will return to the Turbidity Cal Menu.
7.10.4 Grab Calibration - Turbidity
If desired, the turbidity sensor can be calibrated
against the turbidity reading from another instrument.
The analyzer treats the value entered by the user as
though it were the true turbidity of the sample.
Therefore, grab sample calibration changes the sensi-
tivity, it does not apply an offset to the reading.
This screen appears after selecting
Grab calibration.
To calibrate Pulse Flow:
1. Press the
MENU
button
2. Select
Calibrate.
Press ENTER.
3. Select
Sensor 1
or
Sensor 2
corresponding to
Flow. Press ENTER.
4. Select
Pulse Flow.
Press ENTER.
The following screen will appear.
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
7.11 Pulse Flow
7.11.1 Description
A variety of pulse flow sensors can be wired to the Flow signal input board to measure flow volume, total volume
and flow difference (if 2 Flow signal boards are installed). The Model 1056 Flow signal board will support flow
sensors that are self-driven (powered by the rotation of the impeller paddle-wheel).
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED FLOW SENSOR.
THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Grab Cal
Wait for stable
reading
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Grab Cal
Cal Complete
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Grab Cal
Calibration
Error
Press EXIT
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
SN Calibrate?
Pulse Flow
Measure
Sec. Calibration function
Description
Pulse
Flow
7.11.2 K Factor
A constant value representing pulses/Gal of flow
7.11.3 Frequency/Velocity & Pipe
Alternate cal method – requires manual entry of frequency (Hz)
per velocity and Pipe diameter used
7.11.4 In process Calibration
Calibration based on known volume per unit of time
7.11.5 Totalizer Control
User settings to stop, restart and reset total volume meter
TABLE 7-13 Flow Calibration Routines
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...