5.3.1 Purpose
Most liquid analytical measurements (except ORP)
require temperature compensation. The Model 1056
performs temperature compensation automatically by
applying internal temperature correction algorithms.
Temperature correction can also be turned off. If tem-
perature correction is off, the Model 1056 uses the tem-
perature entered by the user in all temperature correc-
tion calculations.
5.3.2 Procedure
Follow the menu screens in Fig. 5.1 to select automatic
or manual temp compensation, set the manual
reference temperature, and to program temperature
units as °C or °F.
5.4.1 Purpose
The Model 1056 accepts inputs from two sensors and
has two analog current outputs. Ranging the outputs
means assigning values to the low (0 or 4 mA) and high
(20 mA) outputs. This section provides a guide for
configuring and ranging the outputs. ALWAYS
CONFIGURE THE OUTPUTS FIRST.
5.4.2 Definitions
1. CURRENT OUTPUTS. The analyzer provides a con-
tinuous output current (4-20 mA or 0-20 mA) directly
proportional to the process variable or temperature.
The low and high current outputs can be set to any
value.
2. ASSIGNING OUTPUTS. Assign a measurement to
Output 1 or Output 2.
3. DAMPEN. Output dampening smooths out noisy
readings. It also increases the response time of the
output. Output dampening does not affect the
response time of the display.
4. MODE. The current output can be made directly
proportional to the displayed value (linear mode) or
directly proportional to the common logarithm of the
displayed value (log mode).
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Temperature
Units: °C
S1 Temp Comp: Auto
S2 Temp Comp: Auto
S1 Manual: +25.0°C
S2 Manual: +25.0ºC
Figure 5-1. Choosing Temp Units and Manual Auto Temp Compensation
5.3 Choosing Temperature Units And Automatic/Manual Temperature Compensation
5.4 Configuring and Ranging The Current Outputs
Instruction Manual
LIQ-MAN-1056
Section 5.0: Programming the Analyzer - Basics
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...