5.9.2 Procedure – Enter Setpoints
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Enter the desired value for the process measurement
or temperature at which to activate an alarm event.
5.9.3 Procedure – Assign Measurement
Under the Alarms Settings menu, this screen will
appear to allow assignment of the alarm relays. select
an alarm assignment. Additional assignment choices
are shown in Figure X-X depending on which meas-
urement board(s) is installed.
5.9.4 Procedure – Set Relay Logic
Under the
Alarms Settings
menu, this screen will
appear to set the alarm logic. Select the desired relay
logic to activate alarms at a High reading or a Low
reading. USP Safety only appears if a contacting con-
ductivity board is installed.
5.9.5 Procedure – Deadband
Under the Alarms Settings menu, this screen will
appear to program the deadband as a measurement
value. Enter the change in the process value needed
after the relay deactivates to return to normal (and
thereby preventing repeated alarm activation).
5.9.6 Procedure – USP Safety
Under the Alarms Settings menu, this screen will
appear to program the USP alarm setting. Enter the
percentage below the limit at which to activate the
alarm.
5.9.7 Procedure – Normal State
The user can define failsafe condition in software by
programming the alarm default state to normally open
or normally closed upon power up. To display this
alarm configuration item, enter the Expert menus by
holding down the EXIT key for 6 seconds while in the
main display mode. Select Yes upon seeing the screen
prompt: “Enable Expert Menu?”
Under the Alarms Settings menu, this screen will
appear to set the normal state of the alarms. Select the
alarm condition that is desired each time the analyzer is
powering up.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Alarm1 S2 Setpoint
+100.0uS/cm
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Alarm
M
Logic:
High
Low
USP
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Alarm1 Deadband
+000.5uS/cm
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Alarm1 USP Safety
+0%
i
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Alarm2 Normal State
Open
Closed
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Alarm
M
Assign:
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
Interval Timer
Fault
Off
Instruction Manual
LIQ-MAN-1056
Section 5.0: Programming the Analyzer - Basics
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...