98
7.8.2 zeroing The Sensor
The following screen will appear during Zero Cal
The following sub-sections provide you with the
initial display screen that appears for each calibration
routine. Use the
flow diagram for Ozone calibration
at the end of Sec. 7 and the live screen prompts to
complete calibration.
The adjacent screen appears after selecting Ozone
calibration:
S1: 1.234 nA
S2: 1.456 nA
S
N
Zero Cal
Zeroing
Wait
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
Zero Cal
In Process Cal
The following screen may appear if In Zero Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
The following screen will appear if In Zero Cal is
successful. The screen will return to the
Amperometric Cal Menu.
S1: 1.234 nA
S2: 1.456 nA
S
N
Zero Cal
Sensor zero failed
Press EXIT
S1: 1.234 nA
S2: 1.456 nA
S
N
Zero Cal
Sensor zero done
If the In Process Cal is successful, the screen will
return to the Cal sub-menu.
The following screen may appear if In Zero Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
7.8.3 In Process Calibration
The following screen will appear after selecting In
Process Cal
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Calibration
Error
Press EXIT
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Wait for stable
reading.
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...