33
Section 5.0
Programming The Analyzer - Basics
5.1 General
Section 5.0 describes the following programming functions:
Changing the measurement type, measurement units and temperature units.
Choose temperature units and manual or automatic temperature compensation mode
Configure and assign values to the current outputs
Set a security code for two levels of security access
Accessing menu functions using a security code
Enabling and disabling Hold mode for current outputs
Choosing the frequency of the AC power (needed for optimum noise rejection)
Resetting all factory defaults, calibration data only, or current output settings only
5.2 Changing Startup Settings
5.2.1 Purpose
To change the measurement type, measurement units, or temperature units that were initially entered in Quick
Start, choose the Reset analyzer function (Sec. 5.9) or access the Program menus for sensor 1 or sensor 2 (Sec.
6.0). The following choices for specific measurement type, measurement units are available for each sensor meas-
urement board.
Signal board
Available measurements
Measurements units
:
pH/ORP (22, 32)
pH, ORP, Redox, Ammonia, Fluoride,
Custom ISE
pH, mV (ORP)
%, ppm, mg/L, ppb,
µ
g/L, (ISE)
Contacting conductivity
(20, 30)
Conductivity, Resistivity, TDS, Salinity,
NaOH (0-12%), HCl (0-15%), Low H2SO4,
High H2SO4, NaCl (0-20%),
Custom Curve
µ
S/cm, mS/cm, S/cm
% (concentration)
Toroidal conductivity
(21, 31)
Conductivity, Resistivity, TDS, Salinity,
NaOH (0-12%), HCl (0-15%), Low H2SO4,
High H2SO4, NaCl (0-20%),
Custom Curve
µ
S/cm, mS/cm, S/cm
% (concentration)
Chlorine
(24, 34)
Free Chlorine, pH Independ. Free Cl, Total
Chlorine, Monochloramine
ppm, mg/L
Oxygen
(25, 35)
Oxygen (ppm), Trace Oxygen (ppb),
Percent Oxygen in gas, Salinity
ppm, mg/L, ppb, µg/L % Sat, Partial
Pressure, % Oxygen In Gas, ppm
Oxygen In Gas
Ozone (-26, -36)
Ozone
ppm, mg/L, ppb,
µ
g/L
Temperature (all)
Temperature
°C. ºF
5.2.2 Procedure
Follow the Reset Analyzer procedure (Sec 5.8) to reconfigure the analyzer to display new measurements or
measurement units. To change the specific measurement or measurement units for each signal board type,
refer to the Program menu for the appropriate measurement (Sec. 6.0).
TABLE 5-1. Measurements and Measurement Units
5.1 General
5.2 Changing Start-Up Settings
5.3 Programming Temperature
5.4 Configuring and Ranging 4-20ma Outputs
5.5 Setting Security Codes
5.6 Security Access
5.7 Using Hold
5.8 Resetting Factory Defaults – Reset Analyzer
5.9 Programming Alarm Relays
Instruction Manual
LIQ-MAN-1056
Section 5.0: Programming the Analyzer - Basics
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...