The following sub-sections show the initial display
screen that appears for each calibration routine. Use
the
flow diagram for Conductivity calibration
at the
end of Sec. 7 and the live screen prompts to complete
calibration.
The adjacent screen appears after selecting
Conductivity calibration:
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
Zero Cal
In Process Cal
Cell K: 1.00000/cm
To calibrate toroidal conductivity:
1. Press the
MENU
button
2. Select
Calibrate
. Press ENTER.
3. Select
Sensor 1
or
Sensor 2
corresponding to
Toroidal Conductivity. Press ENTER.
4. Select
Conductivity
. Press ENTER.
The adjacent screen will appear. To calibrate Toroidal
Conductivity or Temperature, scroll to the desired item
and press ENTER
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibrate?
Conductivity
Temperature
Measure
Sec.
Calibration function: default value Description
Toroidal
Conductivity
7.5.2
Cell K: 3.00000/cm Enter the cell Constant for the sensor
7.5.3
Zero Cal
Zeroing the analyzer with the sensor attached
7.5.4
In Process Cal
Standardizing the sensor to a known conductivity
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
TABLE 7-4
Toroidal Conductivity Calibration
7.5 Toroidal Conductivity Calibration
7.5.1 Description
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instru-
ment, or standardizing to some known reference (such as a conductivity standard). This section contains procedures
for the first time use and for routine calibration of the Model 1056 analyzer.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED INDUC-
TIVE/TOROIDAL CONDUCTIVITy SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...