87
To calibrate free chlorine:
1. Press the
MENU
button
2. Select
Calibrate
. Press ENTER.
3. Select
Sensor 1
or
Sensor 2
corresponding to
Free Chlorine. Press ENTER.
4. Select
Free Chlorine
. Press ENTER.
The adjacent screen will appear. To calibrate Free
Chlorine or Temperature, scroll to the desired item and
press ENTER.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibrate?
Free Chlorine
Temperature
The following sub-sections show the initial display
screen that appears for each calibration routine. Use
the
flow diagram for Chlorine calibration
at the end
of Sec. 7 and the live screen prompts to complete cal-
ibration.
The adjacent screen appears after selecting
Free
Chlorine calibration:
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
Zero Cal
In Process Cal
7.6.1.2 zeroing the sensor
The adjacent screen will appear during Zero Cal. Be
sure sensor has been running in zero solution for at
least two hours before starting zero step.
The adjacent screen will appear if In Zero Cal is
successful. The screen will return to the Amperometric
Cal Menu.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Sensor zero done
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Zeroing
Wait
87
Measure
Sec.
Calibration function: default value Description
Free Chlorine
7.6.1.2 Zero Cal
Zeroing the sensor in solution with zero free chlorine
7.6.1.3
In Process Cal
Standardizing to a sample of known chlorine concentration
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
TABLE 7-5
Free Chlorine Calibration Routines
The adjacent screen may appear if In Zero Cal is unsuc-
cessful. The screen will return to the Amperometric Cal
Menu.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Sensor zero failed
Press EXIT
7.6.1.3 In Process Calibration
The adjacent screen will appear prior to In Process Cal
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Wait for stable
reading.
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...