Section 2.0: Installation
April 2017
FIGURE 2-3. FM Non Incendive Field Wiring Installation
OR
OR
2
2
POWER SUPPLY
ALARM WIRING (VAC) (OPTIONAL)
ANALOG OUTPUT
10
IF A CURRENT LOOP INPUT IS USED WITH CURRENT/FLOW OPTIONS (23/33), THE MAXIMUM ALLOWED VALUES OF THE INPUTS
ARE: 28 VOLTS, 22mA AND 616mW AND THE POWER SOURCE MUST BE FM APPROVED FOR CLASS 1 DIVISION 2.
9. WHEN INSTALLING IN CLASS II AND III LOCATIONS, USE A CLASS II AND III WIRING METHOD
8
MODEL 222, 225, 226, 228, 242 AND 245 TOROIDAL CONDUCTIVITY SENSORS HAVE BEEN APPROVED BY FM FOR USE WITH THE 1056.
ORP SENSORS WITH APPROVED PREAMPS MAY BE USED IN 1056 CLASS I DIVISION 2 INSTALLATIONS. WHEN SELECTED WITH ORP AND
APPROVED PREAMP 23546-00. ORP SENSORS WITHOUT PREAMPS ARE SIMPLE APPARATUS.
SENSORS WHICH ARE FM APPROVED FOR CLASS I DIVISION 2, OR ARE SIMPLE APPARATUS MAY BE USED WITH THE 1056.
SIMPLE APPARATUS IS DEFINED AS AN ELECTRICAL DEVICE THAT DOES NOT GENERATE MORE THAN 1.5V, 100mA, AND 25 mW OR A PASSIVE COMPONENT THAT DOES NOT DISSIPATE MORE THAN 1.3W. CONTACTING CONDUCTIVITY SENSORS AND pH SENSORS WITHOUT PREAMPS MAY BE A SIMPLE APPARATUS, VERIFY THIS WITH THE SENSOR MANUFACTURER.
6
THE EPA AND ISO CLARITY II TURBIDITY SENSORS ARE APPROVED FOR CLASS 1, DIVISION 2
5
4. NO REVISION TO DRAWING WITHOUT FM APPROVAL. THIS IS AN AGENCY CONTROLLED DOCUMENT.
3
GROUND CONNECTION MAY BE MADE IN HAZARDOUS AREA.
2
WHEN CONDUIT IS USED, A SEAL IS REQUIRED AT EACH CONDUIT ENTRANCE.
METAL CONDUIT
METAL CONDUIT
METAL CONDUIT
METAL CONDUIT
5
6
SENSOR CABLE IS SHIELDED
SENSOR CABLE IS SHIELDED
METAL CONDUIT
3
SENSOR 1
8
5
6
CLARITY II TURBIDITY
(OPTIONAL)
CLARITY II TURBIDITY
8
UNCLASSIFIED
AREA
UNCLASSIFIED
AREA
1056
HAZARDOUS
AREA
CLASS I, DIV. 2, GPS A-D, 0° - 50°C
CLASS II, III, DIV 2, GPS E-G
10
10
NOTES: UNLESS OTHERWISE SPECIFIED
SCALE: NONE
WEIGHT:
SIZE
D
DWG NO
SHEET 1 OF 1
1400324
REV
E
AGENCY CONTROLLED DRAWING
ANY CHANGE WILL REQUIRE
CERTIFICATION AGENCY SUBMITTAL / APPROVAL
Содержание Rosemount 1056
Страница 1: ...Instruction Manual LIQ MAN 1056 Rev K April 2017 Rosemount 1056 Dual Input Intelligent Analyzer ...
Страница 4: ...Quick Start Guide Figure A Quick Start Guide ...
Страница 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Страница 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Страница 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Страница 113: ...105 FIGURE 7 1 Calibrate pH Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 114: ...106 FIGURE 7 2 Calibrate ORP Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 117: ...109 FIGURE 7 5 Calibrate Oxygen Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 118: ...110 FIGURE 7 6 Calibrate Ozone Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 119: ...111 FIGURE 7 7 Calibrate Temperature Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 120: ...112 FIGURE 7 8 Calibrate Turbidity Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 121: ...113 FIGURE 7 9 Calibrate Flow Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Страница 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...