CAL.INHIBIT
CAL.INHIBIT is the Inhibit page. Access this screen by pressing the INHIBIT Bu on on the
TOP.CAL page a er pressing the CAL Bu on from the TOP page. This page is for changing the
Inhibit that sets a threshold which the Flow Rate Output must rise above in order for the elec
‐
tronics to begin coun ng. The range for se ng this threshold is from 0.5% of full scale to 4.0%
of full scale.
The Inhibit can be changed by pressing the accompanying Inhibit Bu ons.
Unless OFF is selected for Inhibit, the Total display on the page TOP.RUN and the output signal
‘Frequency, Flow Rate Propor onal’ will be inhibited. The Inhibit for the Flow Rate display on
the page TOP.RUN and the output signal ‘4
‐
20mA’ operate slightly di
ff
erent. The Inhibit is OFF
for the Flow Rate and 4
‐
20mA signal when the 4
‐
20mA moving average
fi
lter is OFF. (See the
page CAL.FILTER.) To Inhibit the Flow Rate and 4
‐
20mA signal, both the 4
‐
20mA moving aver
‐
age
fi
lter and the Inhibit must be on.
INHIBIT BUTTONS
INHIBIT BUTTONS