To successfully u lize the alarm func on of your Digital Electronics Package, please make sure
that the alarm relay is wired according to the diagram in the Wiring Customer Connec ons
‐
Alarm Outputs sec on in Chapter 1 of this manual. The relay can be used to trigger alarms
which will either hold for a period of me a er a set Total has been reached, or will latch in
‐
de
fi
nitely un l the Total is reset. Below is a step by step procedure for se ng a Totaliza on
Alarm.
Step 1:
From the TOP.I/O page press the
ALARM
Bu on. Details of this page
can be found in the HMI Screen
Guide in this manual.
Step 2:
You will now be on the I/O.ALARM
page. Details of this page can be
found in the HMI Screen Guide in
this
manual.
Step 3:
Press the TYPE Bu on. Select
‘WEIGHT’ from the pop up menu.
Check the Blue display box to the
right of the TYPE Bu on to make
sure that “Weight” is displayed.
Step 4:
You have chosen to set the alarm
based on Weight, or Totaliza on.
Now you must select a target value
that will trigger the alarm. Select a
Target Value by
fi
rst pressing the
TARGET
Bu on.
Step 5:
You should now be on the
I/O.ALARM.SP page. Press the
WEIGHT TARGET Bu on. Details of
this page can be found in the HMI
Screen Guide in this manual.
Se ng the Alarm
‐
Totaliza on