56
Phone:(910) 392
‐
2490
Fax: (910) 392
‐
2123
416 Landmark Drive
Wilmington, NC 28412
www.easterninstruments.com
Entering the Digital Electronics IP Address will bring up a web page en tled “G3 Web Server”.
This page will allow you access to three op ons, View Data, View Logs, and Remote View.
Remote View
Selec ng the Remote View op on will allow you to access the various pages of the HMI touch
screen from a remote loca on. Upon clicking the Remote View link, a screen that looks iden
‐
cal to the HMI touch screen will appear in the Web Browser window. Clicking on the various
op ons works similarly to using the touch screen on the Digital Electronics itself. In this way,
all of the features of the HMI can be accessed and used remotely.
View Logs
Selec ng the View Logs op on allows access to the log
fi
les stored on the Compact Flash.
Three
fi
les are created when the compact
fl
ash is forma ed— MMDDYY.csv, UserSet.csv, and
MTR.csv. They are all comma separated variable
fi
les. The MMDDYY
fi
le store the run data.
One run data history
fi
le is created upon startup of the Digital Electronics and a new
fi
le is sub
‐
sequently created every 24 hour period. The names of the
fi
les themselves indicate the date
upon which the
fi
le was created and the data therein was collected. The UserSet
fi
le records a
history of changes to the meter calibra on. The MTR
fi
le records the site calibra on data.
MTR is an abbrevia on for Material Test Report.