Step 6:
Hang the Test Weight provided
with the Electronics Enclosure from
the Calibra on Stud located under
the bo om edge of the Pan as seen
below. For Pans that have more
than one Calibra on Stud, there
should be a Test Weight for each.
Some op onal pans may not have a
Stud. In these cases the wire will
be long enough to hook onto the
top edge of the pan and be draped
over the front end of the pan as
shown below. Be sure that the
Test Weight hangs freely from the
Calibra on Stud. It is also good prac ce to allow the Test Weight to become stable
and movement free before proceeding.
Step 7:
Press the CALs Bu on located on the lower le side of the screen. Note the % Volts
reading displayed in the Diagnos c Window. Verify the reading is correct for the
selected range. For example, 25% is the correct value for the Standard Range se ng.
Please see the table below for the correct % Volts reading for each range.
Step 8:
Remove the Test Weight from the Calibra on Stud. The voltage should return to
0.00%
Volts.
PAN WITH PAN LINER
AND CALIBRATION STUD
ALL
‐
WELDED PAN LINER
WITHOUT A CALIBRATION STUD