8.6 Weld Geometry Display In PA Mode
8.6.1
Weld Geometry Selection
When PA Mode is ON (see 8.1. Initializing
PA Mode), activate the
S-scan
display.
Then depress the left (yellow) button for
selection of weld geometry to be scanned.
The MODE WELD OFF changes to MODE
WELD GEOM; at the same time, the
S-scan
changes to
WELD
.
There are some predefined examples of
weld geometry that can be chosen by the
WELD X of N buttons or can be designed
“on the fly” – see below.
Adjust the PROBE – WELD parameter
(distance from the Wedge Front to the weld
axis) for correct displaying of weld position
relative to the PA probe.
It may be needed to correct the THICKNESS value as well, for displaying the correct
weld geometry. This parameter is the same as can be seen in the
SETTING
menu.