Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
39/102
BSCAN Menu Items on the Screen – Right Side:
4.5.1.6 GAIN
Adjustment of the input gain value – 0.0 dB to 111.0 dB. See also NOTES in 4.2.6.
4.5.1.7 BSCAN MAX
Adjustment of maximum limit thickness value.
4.5.1.8 BSCAN MIN
Adjustment of minimum limit thickness value.
4.5.1.9 THICKNESS
-
Adjustment of minimum thickness value for automated thickness measurement using
Gate 2 – this value must be lower than any really expected thickness to be measured.
-
Adjustment of nominal thickness line in the B-scan thickness measuring mode.
4.5.1.10 VELOCITY
Adjustment of the sound velocity value in the tested material – 1 m/s to 19999 m/s.
For details see 4.1.11. VELOCITY.