4.5.3.2 Automatic thickness measuring mode with GATE 2 set to THICKNESS
AUTO 1P-1P:
1. Set GATE 1 over at least two
consecutive back wall echoes.
2. Switch GATE 2 mode to THICKNESS
AUTO 1P-1P (from 1
st
peak within
GATE 1 to 1
st
peak within GATE 2).
3. Proper minimum THICKNESS shall
be set in GATES menu.
4. Activate BSCAN menu,
-
BSCAN MAX value shall be set to
the top thickness limit of the graph
(mm or inch);
-
BSCAN MIN value shall be set to
the bottom thickness limit of the
graph (mm or inch).
4.5.3.3 Thickness measuring mode with GATE 2 set to THICKNESS 1P-1P:
1. Set GATE 1 over first back wall echo.
2. Switch GATE 2 mode to THICKNESS
1P-1P (from 1
st
peak within GATE 1 to
1
st
peak within GATE 2).
3. Set GATE 2 over second back wall
echo.
4. Activate BSCAN menu,
-
BSCAN MAX value shall be set to the top thickness limit of the graph (mm
or inch);
-
BSCAN MIN value shall be set to the bottom thickness limit of the graph
(mm or inch).
4.5.4
R-G-B Palette
After R-G-B option had been chosen in the
SETTING
→
DISPLAY
sub-menu
(see 4.7.5 in 4.7. DISPLAY Sub-Menu (SETTING)), depressing the BSCAN key
displays rolling colour graphic presentation of echo amplitudes with colours
corresponding to the colour palette shown at the right border of the display.