Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
65/102
7. Applications
7.1 Automatic Thickness Measurement
See also 5.2.2 for GATE 2 automatic thickness measurement function.
1. Adjust appropriate VELOCITY and
Time Base.
2. Switch to the GATES menu.
3. Adjust
minimum
THICKNESS
value. The minimum THICKNESS
value
shall
be
less
than
expected thickness value to be
measured.
4. Place transducer on the subject to
be measured and get a waveform with at least two consecutive back-wall
echoes.
5. Adjust GATE 1
- vertical level to cross both echoes to be measured,
- horizontal position and width so as to cover both echoes to be measured.
6. Switch GATE 2 to the THICKNESS AUTO 1P-1P mode
7. The measured thickness value will be displayed in the Status window with
dark blue colour.