4.1.2
GATE --<
Adjustment of the gate end position (changes
its width) – actual gate number (activated in
the
GATES
menu – see 4.4) is indicated in
parentheses next to the gate end position
value.
4.1.3
GATE ^
Adjustment of the gate vertical position – actual
gate number (activated in the
GATES
menu – see 4.4) is indicated in parentheses next
to the gate vertical position value.
4.1.4
TR MODE – Transmitter/Receiver Mode
Switches the Transmitter/Receiver modes as follows:
-
within ENHANCED mode (
HW_SET
with yellow background), both left and right
buttons next to TR MODE switch to the BASIC mode (the message
BACK TO
BASIC MODE ? YES CANCEL
will appear);
-
within BASIC mode (
ENHANC.
with grey background instead
HW_SET
), the
following Transmitter/Receiver modes are available:
-
SINGLE – for pulse-echo technique with single element probe,
-
DUAL – for through-transmission technique (two separate probes) or pulse-
echo technique with dual element probe (TR probe),
-
EMAT – for pulse-echo technique using EMAT probe (Electromagnetic
Acoustic Transducer);
NOTE: After switching to the BASIC mode, the parameters in the
HW_SET
menu
(see 4.2) will be set to default values as follows:
-
BANDWIDTH – full,
-
ATT – auto,
-
DAMP – OFF,
-
PRF – 200 Hz auto,
-
Pulser – SPIKE 1.
-
within ENHANCED mode (
HW_SET
with green background), following
Transmitter/Receiver modes are available:
-
SINGLE – for pulse-echo technique with single element probe,
-
DUAL – for through-transmission technique (two separate probes) or pulse-
echo technique with dual element probe (TR probe),
-
DUAL+ExtAmp – for through-transmission technique (two separate probes)
or pulse-echo technique with dual element probe (TR probe), where receiving
probe (element) is equipped with built-in preamplifier,
-
EMAT – for pulse-echo technique using EMAT probe (Electromagnetic
Acoustic Transducer).