Supplement Scio Four modules
Monitoring
Set up and use the patient monitor in accordance with its instruc-
tions for use.
Tasks before patient change
1
Clean and disinfect the gas analyzer according to hospital
hygiene regulations.
2
Check the water trap on the front of the gas analyzer. Empty
or replace the water trap as needed.
3
When no bacterial filter is used between sample line and
patient, replace the sample line. Dispose of used sample lines.
Switching off the gas analyzer
1
Set the On/Off switch on the rear of the device to "OFF"
Cleaning and disinfection
Cleaning of the gas analyzer
Classification of medical devices
For cleaning, the medical devices and their components are clas-
sified according to their type of application and the resulting risks:
– Non-critical medical devices: surfaces accessible to the user
and patient, e.g., device surfaces, cables
Testing of procedures and agents
Cleaning and disinfection were tested using the following proce-
dures and agents. The following agents showed good material
compatibility and effectiveness at the time of the test:
– Dismozon plus from the company Bode Chemie
When selecting a suitable disinfectant, adhere to the country-spe-
cific lists of disinfectants. The list of the German Association for
Applied Hygiene (Verbund für Angewandte Hygiene VAH) applies
in German-speaking countries.
Strictly observe the manufacturer's information on the disinfec-
tants. Manufacturers may change the composition of disinfectants
over time.
Manual disinfection and simultaneous cleaning
At the time of validation, the listed surface disinfectants showed
good material compatibility. Dräger recommends using a surface
disinfectant from the following list.
The manufacturers of the surface disinfectants have verified at
least the following spectra of activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Observe the specifications of the surface disinfectant manufactur-
ers. Other surface disinfectants are used at own risk.
WARNING
Risk of infection
Used sampling tubing, HMEs, water traps, and the device
itself could be contaminated and must be handled and dis-
posed of with care.
Dispose of the accessories in an adequate way and in com-
pliance with local regulations.
WARNING
Risk of infection
Use only validated cleaning and disinfection procedures for
the device and accessories as described in these instruc-
tions for use.
WARNING
Risk due to inaccurate gas measurement values and gas
measurement failure
To avoid temporary influence on the gas measurement and
prevent damage to the water trap and measuring system do
not wash or disinfect the inside of the sample line or water
trap. Do not sterilize the sample line or water trap.
CAUTION
Risk of gas measurement failure
To prevent damage to the gas measurement, especially during
cleaning and disinfection, always keep the water trap installed.
CAUTION
Risk of damaging the pneumatic system
Do not apply excessive pressure (e.g., syringe, or compressed
air) to the inlet, exhaust port, or the water trap of the medical
device, e.g., during cleaning and disinfection.
Class of active ingre-
dient
Product name
Manufacturer
Chlorine-releasing
agents
Actichlor
®
plus
Ecolab
BruTab 6S
®
Brulin
Clorox Profes-
sional Disinfecting
Bleach Cleaner
Clorox
Dispatch
®
Hospi-
tal Disinfectant
Towels with
Bleach
Klorsept
®
17
Medentech
Oxygen-releasing
agents
Descogen
®
Liquid Antiseptica
Descogen
®
Liquid
r.f.u.
Dismozon
®
plus
Bode Chemie
Dismozon
®
pur
Oxycide
®
Ecolab USA
Perform
®
Schülke & Mayr
Virkon
®
DuPont
Quarternary ammo-
nium compounds
Mikrozid
®
sensitiv
liquid
1)
1) Virucidal against enveloped viruses
Schülke & Mayr
Mikrozid
®
alcohol
free liquid
Mikrozid
®
sensi-
tive wipes
Mikrozid
®
alcohol
free wipes
acryl-des
®
Aldehydes
Buraton
®
10 F
Содержание Scio Four
Страница 19: ...Supplement Scio Four modules 19 This page intentionally left blank...
Страница 23: ...Scio Four 23 MC IEC 60601 1 2 MC MC MC MEDIBUS 0 3 1 0 MEDIBUS 38 Dr ger...
Страница 27: ...Scio Four 27 1 RS232C X2 A 2 Infinity 1 HME Luer Lock 2 1 2 002 A...
Страница 28: ...28 Scio Four 1 2 1 2 3 1 1 2 HME...
Страница 30: ...30 Scio Four 1 2 3 4 1 2 3 4 Dr ger Dr gerService 1 1 12 10...
Страница 39: ...Scio Four 39...
Страница 129: ...Dodatak Scio Four moduli 129 Ova je stranica namjerno ostavljena prazna...
Страница 151: ...Scio Four 151 MEDIBUS Scio Four CO2 N2O O2 0 3 m 1 0 MEDIBUS 166 Dr ger PVC...
Страница 152: ...152 Scio Four 25 MRI NMR NMI 25 cm 10 Dr ger 001 A B C 002 A B C D E F G MEDIBUS H RS232C X2 A B C A B C E G H F D...
Страница 155: ...Scio Four 155 1 Y Luer Lock 2 1 2 AGSS 1 2 O 1 2...
Страница 156: ...156 Scio Four 1 2 3 1 Dismozon plus Bode Chemie...
Страница 163: ...Scio Four 163 CISPR 11 1 A IEC 61000 3 2 IEC 61000 3 3...
Страница 167: ...Scio Four 167...
Страница 171: ...Scio Four 171 MEDIBUS Scio Four CO2 N2O O2 0 3 m 1 0 ft MEDIBUS 186 Dr ger PVC...
Страница 172: ...172 Scio Four 25 MRI NMR NMI 25 cm 10 in Dr ger 001 B 002 B MEDIBUS H RS232C X2 B B H...
Страница 175: ...Scio Four 175 1 Luer Lock 2 1 2 1 2...
Страница 176: ...176 Scio Four 1 2 1 2 3 1...
Страница 178: ...178 Scio Four 1 2 3 IEC 62353 4 Scio Four Scio Four Oxi 2 Vol 0 50 Vol Dr ger Dr gerService 1 1 12 10 min 10 200 mA DC...
Страница 187: ...Scio Four 187...
Страница 205: ...Ek Scio Four mod lleri 205 Bu sayfa kasten bo b rak lm t r...
Страница 209: ...Scio Four 209 Scio Four O2 CO2 N2O Dr ger IV PVC 25 MRI NMR NMI 25cm 10in Dr ger 001 A B C A B C...
Страница 212: ...212 Scio Four 1 Y HME Luer 2 1 2 1 On Off ON 2 O 1 2...
Страница 225: ...Scio Four 225 MEDIBUS Scio Four CO2 N2O O2 MEDIBUS 237 Dr ger IV PVC PVC 25 MRI NMR NMI 25cm 10 Dr ger...
Страница 229: ...Scio Four 229 1 ON OFF ON 2 1 2 3 OFF 1 ON OFF OFF 1 2 HME...
Страница 239: ...Scio Four 239...