PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 80
4.3.2.1
ENHANCED PROGRAM MESSAGE USING THE COMMAND TYPE
The COMMAND type of program message executes a process and can additionally
send data to the PPC2+ in the form of comma delimited arguments. This data is
usually a setting of some sort that is stored in the PPC2+. If data is specified, it must
be preceded by at least one white space from the program message header and be
within the range and format described in the program message description.
The
keyword “Command:” will appear to the left of the required syntax in each
program message description.
If you are using the IEEE-488 port, the Command type does not generate a reply so
you do not need to wait for a reply. You also may send multiple program messages at
once by separating each program message with a semicolon. The commands will be
queued and executed in as received order.
If you are using the RS-232 port COM1, the Command type will always generate a
reply so you must wait for a reply before issuing another program message.
Because of this, you can only send one Command program message at a time while
using the COM1 port.
4.3.2.2
ENHANCED PROGRAM MESSAGE USING THE QUERY TYPE
The Query type of program message just requests data from the PPC2+. You must
wait for a reply with a query. If you send any type of program message to the PPC2+
after a Query before receiving a reply, the program message will be discarded and an
error will be generated. Errors are reporting using the IEEE Std. 488.2 status
reporting model. A Query program message always ends with a question mark.
Most queries will return a reply within 200 ms except:
PR?, PRR?, SR?, ATM?, RATE? Up to 2.0 seconds to allow a new measurement.
The syntax for using a QUERY program message is listed next to the keyword
“Query:” in each program message summary in Section 4.4.4.
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...