PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 122
∗
OPC
Purpose
Sets the operation complete bit when all operations have completed.
Command
Query
“
∗
OPC”
“
∗
OPC?”
Remarks
This Command enables the PPC2+ to set the OPC bit in the Standard Event Status
Register when it has completed all pending functions.
The Query replies with a “1” when all functions are complete.
Example
(enhanced)
Sent:
Query reply:
“
∗
OPC”
“1”
∗
OPT
Purpose
Reads the list of installed PPC2+ options.
Query “
∗
OPT?”
Remarks
This Query returns any registered option(s) installed in the PPC2+. Each option is
separated by a comma. Possible options:
“IEEE-488:0”
“ANALOG:n”
The IEEE-488 option is installed.
The analog option is installed. ‘n’ is the revision on the analog
option hardware from ‘A’ to ‘Z’ or ‘-’ if original revision.
Example
(enhanced)
Sent:
Reply:
“
∗
OPT?”
“IEEE-488:0, ANALOG:-”
∗
RST
Purpose
Resets the PPC2+ control settings to factory settings.
Command “
∗
RST””
Remarks
This Command sets the PPC2+ settings to factory settings. This equivalent to a
front panel executed RESET/SET. This does not affect the communications
settings.
Example
(enhanced)
Sent:
Reply:
“
∗
RST”
“
∗
RST” (no reply if IEEE-488)
See Also
Section 3.4.4.5, Reset - Sets
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...