PPC2+™ Operation and Maintenance Manual
Page 133
©2001 DH Instruments, Inc.
ñ
Enter the calibration pressure and PPC2+ readings into a spreadsheet. Calculate the
“non-corrected” PPC2+ readings by backing out the PA, PM and ZOFFSET (absolute
RPTs only) recorded in Steps
ê
and
above, following:
non-corrected reading = (corrected reading - PA + ZOFFSET)/PM
ò
Perform a linear regression to find the offset and slope that best fit the non-corrected
PPC2+ readings to the calibration standard pressures. The offset is the new value of
PA, the slope is the new value of PM.
ô
Press
[SPECIAL] and select <4Internal>, <2Cal>, <1RPT>, <2edit> and write the
new calibration date and the new values of PA/PM for the RPT and range calibrated.
õ
Press
[SPECIAL] and select <4Internal>, <2Cal>, <1RPT>, <3run ZNATERR>, run
the ZNATERR routine (absolute RPTs only) (see Section 5.2.6).
ù
Calculate
as
left data for the calibration if desired following:
as left reading = (non-corrected reading o new PM) + new PA
5 . 3 A D J U S T M E N T O F O N - B O A R D B A R O M E T E R
m
PURPOSE
To adjust the output of the on-board barometer (see Section 1.2.2.2).
m
PRINCIPLE
The on-board barometer output can be adjusted using PA and PM values in the same manner as
for the reference pressure transducers (see Section 5.2.1).
>ROYXLYK\NLK\YWO^O\S]KVYaKMM_\KMc]OX]Y\_]ONYXVcPY\WOK]_\SXQMRKXQO]SXK^WY]ZRO\SM
Z\O]]_\OY`O\]RY\^ZO\SYN]YP^SWO]OO=OM^SYX::-WOK]_\OWOX^KMM_\KMcNYO]XY^
NOZOXNYX^ROKL]YV_^OKMM_\KMcYP^ROYXLYK\NLK\YWO^O\
m
OPERATION
To edit the values of PA and PM for the barometer, press [SPECIAL] and select <4Internal>,
<2cal>, <2barometer>. Pressing [ENTER] steps through displays of the calibration date
[YYMMDD] and PA/PM. In edit mode, the values can be edited. Pressing [ENTER] after the last
screen activates the edited values.
>Y `SOa ^RO M_\\OX^ K^WY]ZRO\SM Z\O]]_\O WOK]_\OWOX^ WKNO Lc ^RO LK\YWO^O\ LYK\N Z\O]]
E=:/-3+6GKXN]OVOM^&3X^O\XKV(&!+^W(
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...