PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 120
4.5.4
READY STATUS REGISTER
The PPC2+ contains an 8 bit Ready Status Register that reflects specific PPC2+
measurement and generation ready events. Enabled events in this register will set or clear
the RSB bit of the Status Byte Register.
Table 16. 8 Bit Ready Status Register
N/A
(128)
N/A
(64)
N/A
(32)
N/A
(16)
N/A
(8)
MEAS
(4)
NRDY
(2)
RDY
(1)
This register can be read using the “*RSR?” query, Each of these status bits can set the RSB
bit of the Status Byte Register, causing a SRQ to occur IF the RSB bit is enabled to do so.
The Standard Event Status Enable Register (“
∗
RSE” program message ) determines which of
these flags are able to assert the RSB bit. The description of these bits are given as:
•
MEAS: Measurement ready (Bit 2)
Indicates that the PPC2+ has completed an RPT measurement.
•
NRDY: Generation Not Ready (Bit 1)
Indicates that the PPC2+ made a transition from Ready (<*>) to Not Ready (<
↑
> or <
↓
>)
as defined by the control settings (see Section 3.1.2.4).
•
RDY: Generation Ready (Bit 0)
Indicates that the PPC2+ has reached a target pressure and is Ready (<*>) as defined by
the control settings (see Sections 3.1.2.4 and 3.1.2.3).
4 . 6 I E E E S T D . 4 8 8 . 2 C O M M O N A N D S T A T U S P R O G R A M
M E S S A G E S
The PPC2+ supports a set of commands that are common to all instruments conforming to
IEEE Std. 488.2. These command make it easy to perform basic function for any device that supports
these commands. These command also cover the status reporting commands. See Section 4.5 for
details on the status registers mentioned in these commands.
Table 17. Program Message List
∗
CLS
Clear all of the status & event structures.
∗
ESE
Read or set the Standard Event Status Enable Register.
∗
ESR
Read the Standard Event Status Register.
∗
IDN
Identify the PPC2+ version, range, and serial number.
∗
OPC
Sets the operation complete bit when all operations have completed.
∗
OPT
Reads the list of installed PPC2+ options.
∗
RST
Resets the PPC2+ control settings to factory settings.
∗
TST
Read the power on self test status.
∗
SRE
Read or set the Service Request Enable Register.
∗
STB
Read the Status Byte Register.
RSE
Read or set the Ready Status Enable Register.
RSR
Reads the Ready Status Register.
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...