PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 106
RES
Purpose
To read or set the pressure display resolution for the current RPT and range.
Command
Query
“RES
n”
“RES”
Classic “RES=
n”
“RES”
Default “RES
0.001”
Arguments
n:
The pressure display resolution in % FS of the current RPT range
(0.0001 to 1 % FS).
Remarks
The pressure display resolution is defined as % FS of the selected RPT range. The
setting is separate for each range, and will change as the range is changed.
Example
(enhanced)
Cmd sent:
Query reply:
“RES .01”
“0.01 %FS”
Example
(classic)
Sent:
Reply:
“RES=.01”
“0.01 %FS”
Errors
ERR# 6
The argument is invalid.
See
Also Section
3.3.2
RESET
Purpose
Reset the user’s settings to factory defaults.
Command “RESET”
Classic “RESET”
Remarks
The PPC2+ has user settings (units, resolution, control modes, etc.) that can be
reset to factory defaults. Calibration sequences and communications settings will
not be affected. The remote “RESET” program message corresponds to the local
“Reset-sets.” function (See 3.4.4.5, Reset-sets.) The reset cycle takes up to 10
seconds to complete. Remote communications should not take place during this
period.
Example
(enhanced)
Cmd sent:
Reply:
“RESET”
“RESET” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“RESET”
“RESET”
See Also
Section 3.4.4.5, Reset-sets
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...